johnodonahue Publish time 2019-12-29 11:09:53

Hantek DSO5102P Error Code 0xc00

Scope powers up and appears to work, but waveform is off scale at top using 5V@1KHz reference.Rotating vertical position knob makes trace visible briefly.Measurement display showing 94 volts.Using separate 4MHz oscillator shows similar result.5V@1KHz reference signal looks correct on older analog oscilloscope.Self Calibrate gives Error Code 0xc00 at step 2/38


wangshuang Publish time 2019-12-31 08:06:34

Please send a photo of the product information in Utility to my e-mail: wangshuang@hantek.com

luis67 Publish time 2020-8-11 00:40:49

wangshuang replied at 2019-12-31 08:06
Please send a photo of the product information in Utility to my e-mail:

Hi. I have the same problem with my Hantek DSO5202P, have you managed to solve it? regards

wangshuang Publish time 2020-8-12 11:26:14




Replace the relay and 8370.

ylkhong1@gmail.com Publish time 2022-4-4 20:21:18

Anyone tried rwplacement ro confirm it works? I have samw problem