ying Publish time 2020-6-30 18:08:53

Hantek 1833C Calibration

Customer Consultation


Hello, This week I purchased your LCR meter Hantek 1833C and happy with the measurement result. Unfortunately I cannot find in the Hantek 1833C manual how to make the calibration. Could you please clarify? Thanks,

ying Publish time 2020-6-30 18:10:12

Edited by ying at 2020-7-2 10:46

Reply

Please download the manual from the website: http://hantek.com.cn/products/detail/16180 And refer to chapter 2.6 Correction function.

Russ3000 Publish time 2021-6-5 16:42:29

I really like your 1832c, its beautiful design, accurate measurements and affordable price, so why the hell does it not remember the calibration results after turning off the power? Please pass it on to the developer. Let's make the device better together!

ying Publish time 2021-6-22 16:26:27

Russ3000 replied at 2021-6-5 16:42
I really like your 1832c, its beautiful design, accurate measurements and affordable price, so why t ...

Due to the drift of the instrument itself, adding some stray elements, such as contact resistance, inductance, shunt capacitance, etc.. In order to guarantee the accuracy of the instrument, remove measuring fixture or wire and test the stray capacitance and inductance and lead resistance influence on measurement accuracy, must to reset instrument, instrument reset including two reset open circuit and short circuit.

Redy Publish time 2022-5-31 03:59:02

Hi Yin,
after calibrate it's OK, but when I turn OFF device and turn ON again ... my C value always at 8pf .... can't going to 0 pf, any suggestion till result calibration can save for future use?

vedihe20 Publish time 2023-8-5 20:11:36

Calibration is a critical process when it comes to ensuring the accuracy and reliability of your measuring instruments. The Hantek 1833C oscilloscope is no exception.