Koloholm Publish time 2021-5-20 00:07:25

Hantek DSO 2D10 the frequency counter shows incorrectly

Edited by Koloholm at 2021-5-20 03:48

Hantek DSO 2D10 the frequency counter shows incorrectly, the built-in frequency counter shows the signal frequency 2 times less than the real one, for example, if you connect the oscilloscope input to a calibrator or generator, it shows a frequency of 500 Hz, although the frequency of the generator is 1 kHz, and so on at all frequencies.




amy Publish time 2021-5-20 08:47:10

Edited by amy at 2021-5-20 16:29

Please download the latest firmware to update.
http://hantek.com/products/detail/17182

Koloholm Publish time 2021-5-20 13:21:16

what kind of firmware you need to specify so that there are no glitches in it later

amy Publish time 2021-5-20 16:30:15

There is only one latest firmware file in the link.

Koloholm Publish time 2021-5-20 20:07:14

Edited by Koloholm at 2021-5-21 16:11

Many users of these oscilloscopes complain that other glitches appear after flashing.

Koloholm Publish time 2021-5-22 22:04:11

after flashing it worked fine