WisdomAugust Publish time 2023-4-3 08:39:07

DSO6000 sereis guide is based on LabView8.2.

General Settings


1 Voltage


As shown in the figure, changing the number of the red square can change the voltage
scale value range from 0 to 11. For details, please refer to Table 3 in "SDK_HTHardDll_EN".
For example, 4864 indicates that CH1 is set to 50mV, CH2 1V, CH3 200mV, CH4 50mV.






WisdomAugust Publish time 2023-4-4 08:16:41

Table 3 Range of each voltage base




Note:
1 "Voltage division" means a large vertical grid voltage corresponding to the value, more precisely, is a
waveform data acquisition data 32 corresponding to the difference value 2 "Range" is represented by
1: 1 probe corresponding to the range. For example, 100mV with a 1: 1 scale probe is 800mV; 1:10
probe range is 8V



WisdomAugust Publish time 2023-4-6 08:29:54

2 Channel Switch




0 means off, 1 means on, in the figure example
all 4 channels are on.



WisdomAugust Publish time 2023-4-7 08:15:55

3 Timebase/Sampling Rate



Just change the numbers in the picture as shown. For
details, refer to Table 1 in "SDK_HTHardDll_EN".
In the figure example, 12 indicates that the time base
is 20uS and the sampling rate is 12.5MSa.




WisdomAugust Publish time 2023-4-8 08:40:16

Table 1 Sampling Rate





WisdomAugust Publish time 2023-4-10 08:16:06

Note:


1 Column " Sampling rate " shown in bold need Interpolation.
2 When the oscilloscope does not require interpolation, sample rate = 250 /timebase;
250 is the number of points in a single grid, such as when the timebase is 1uS sampling


rate = 250/(1e-6) = 250MSa


3 "S" indicates single-channel mode; "D" means dual-channel mode; "Q" means
four-channel mode. About channel mode is shown in Table 2: timebase and channel mode.


WisdomAugust Publish time 2023-5-4 13:12:32

4 Bandwidth Limit/Input Coupling


a. 20M bandwidth limit: 0 means off 1 means on, the red rectangle represents the bandwidth limit of 4
channels.


b. Input coupling: 0 means AC, 1 means DC, and the yellow bottom word means the input coupling of
4 channels.






WisdomAugust Publish time 2023-5-5 08:12:14

5 Vertical Trigger Position


The yellow part as shown in the figure indicates the value of the vertical trigger position.
The range is 0-255.
The example in the figure is 191.





WisdomAugust Publish time 2023-5-8 13:08:47

6 Horizontal Trigger Position


The yellow part as shown in the figure indicates the value of the horizontal trigger position.
The range is 0-100.

In the figure example is 50.





WisdomAugust Publish time 2023-5-9 08:18:07

Remarks:
For all Dll functions of LabView, please refer to the file SDK_HTHardDll_EN.pdf. For example, if you don't
know the usage of, you can double-click to get the prototype as "dsoHTGetState", and then you
can get the meaning of each parameter in SDK_HTHardDll_EN.pdf.


Pages: [1]
View full version: DSO6000 sereis guide is based on LabView8.2.